Routine shock-energy testing in primary-prevention ICDs questioned—again
August 4, 2008 | Steve Stiles

Washington, DC - A retrospective look at SCD-HeFT has added to a good deal of observational evidence questioning the routine need for defibrillation threshold (DFT) testing in most patients receiving implantable cardioverter defibrillators (ICDs) for primary prevention of sudden death [1].

Advances in ICD technology and experience "have called into question whether DFT testing at the time of implantation is still required. Yet physicians have been reluctant to abandon it, despite mounting evidence that it may no longer be necessary," writes Dr Anne B Curtis (University of South Florida, Tampa) in an editorial [2] accompanying the SCD-HeFT analysis in the August 12, 2008 Journal of the American College of Cardiology.

The new SCD-HeFT analysis suggests that DFT testing had little bearing on the efficacy of delivered shocks or long-term mortality. It supports other observational studies suggesting that it offers little clinical value in most stable patients receiving primary-prevention ICDs who don't have an indication for resynchronization therapy but who were on "standard-of-care background medical therapy for heart failure," according to Curtis, a recent president of the Heart Rhythm Society.

"Proof of concept has now been established that routine DFT testing is unlikely to provide sufficient information that impacts ICD programming in the overwhelming majority of patients, and it carries some slight risk," she concludes.

The current analysis doesn't, on its own, prove that DFT testing shouldn't be performed in primary-prevention ICD recipients, "but when we look carefully at the data, we don't really see that it provides any particular benefit," Dr Joseph A Blatt (University of Washington, Seattle), lead author of the analysis, told heartwire.

As we get better at device implantation, we have to think about whether putting the patient under further risk is really worth it.

"As we get better at device implantation, we have to think about whether putting the patient under further risk is really worth it. We know that there is a chance of sudden death after DFT testing, although it's low, in the range of 0.1% or less," Blatt said.

Embolic strokes have also been described as a complication, he observed, and there is increasing evidence that the shocks themselves can somehow damage the myocardium. "Future studies are needed to really tease this out and to see whether or not [DFT testing] is beneficial or possibly harmful."

SCD-HeFT randomized more than 2500 patients with NYHA class 2-3 heart failure and LVEF <35% but no history of sustained VT or cardiac arrest to receive ICDs, amiodarone, or placebo on top of then-standard therapies. Those getting devices received single-lead Medtronic ICDs programmed to deliver only shock therapy with a maximum output of 30 J. Importantly, antitachycardia pacing wasn't a treatment option.

The findings of a 23% reduction in mortality (p=0.007) over five years, regardless of HF etiology, and no significant effect in the amiodarone group were publicized in March 2004 and published the following year [3] and have been extensively covered by heartwire.

The analysis from Blatt et al included the trial's 811 patients who had been randomized and agreed to receive ICDs; baseline DFT thresholds were known for 717. The average follow-up was almost four years. SCD-HeFT had excluded patients with an indication for dual-chamber pacing and forbade the use of biventricular devices.

The success rate for the first appropriate shock ranged nonsignificantly from 81.0% to 100% among patients with implant DFTs of <10 J, 11-20 J, and 21-30 J and those with unknown baseline DFTs. The implantation DFT was <20 J in all but three, or 97.8%, of the 717 for whom it was known.

Mortality was 20.7% among the 547 patients with DFTs <10 J and 19.4% for those with higher DFTs, for an adjusted hazard ratio of 1.19 (p=0.41) for the latter group; the results were similar whether the 94 patients without baseline DFT data were included in the lower-DFT or higher-DFT group.

Blatt said the SCD-HeFT outcomes are entirely relevant to device therapy today. Despite the trial's exclusive use of single-lead defibrillators limited to therapy with shocks, the devices were "essentially modern" in terms of implantation technique, shock waveforms, and levels of delivered energy. Concomitant HF drug therapy was also consistent with today's standards, he said.



"Strategic" programming reduces ICD-related morbidity

A programming strategy for primary-prevention ICDs aimed at optimizing the use of antitachycardia pacing (ATP) and reducing the burden of appropriate and inappropriate shocks, compared with individually tailored device programming, can lower overall device-related morbidity, conclude investigators from the prospective Primary Prevention Parameters Evaluation (PREPARE) [4].

"By moderately extending the detection duration, using fast rate cutoffs, ATP in faster rate zones, and existing [supraventricular tachycardia] discrimination algorithms, the PREPARE-study programming can safely be used to reduce the combined morbidity of spontaneous shocked episodes, arrhythmic syncope, and untreated sustained symptomatic VT/VF events in primary-prevention patients," write Dr Bruce L Wilkoff (Cleveland Clinic Foundation, OH) and associates in the August 12, 2008 Journal of the American College of Cardiology.

The study had been reported in preliminary form at the May 2007 scientific session of the Heart Rhythm Society and covered by heartwire at the time.

The PREPARE programming was associated with significant improvement, from 0.69 to 0.26 events per patient-year (p<0.003), in an index based on the three morbidity end points (spontaneous shocks, arrhythmic syncope, and sustained symptomatic VT/VF) in 700 patients receiving primary-prevention single- or dual-chamber or biventricular ICDs. The cohort also showed a 0.49 hazard ratio for adjusted risk of shock from any cause during the first follow-up year (95% CI 0.29-0.82, p<0.01).

The comparator group consisted of 689 similar primary-prevention ICD recipients from the Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter Defibrillators (EMPIRIC) and Multicenter InSync Implantable Cardioversion Defibrillation Randomized Clinical Evaluation (MIRACLE-ICD) trials, in which device programming was left to physician discretion.

—SS


SCD-HeFT was supported by the US National Institutes of Health, Medtronic, and Wyeth-Ayerst Laboratories. Curtis reports consulting for and receiving honoraria and grant support from Medtronic, receiving honoraria from Boston Scientific, and consulting for and receiving honoraria from St Jude Medical. PREPARE was sponsored by Medtronic. Disclosures for specific SCD-HeFT and PREPARE coauthors are included in the papers.

Sources
  1. Blatt JA, Poole JE, Johnson GW, et al. No benefit from defibrillation-threshold testing in the SCD-HeFT (Sudden Cardiac Death in Heart Failure Trial). J Am Coll Cardiol 2008; 52:551-556.
  2. Curtis AB. Defibrillation threshold testing in implantable cardioverter-defibrillators. Might less be more than enough? J Am Coll Cardiol 2008; 52:557-558.
  3. Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005; 352:225-237.
  4. Wilkoff BL, Williamson BD, Stern RS, et al. Strategic programming of detection and therapy parameters in implantable cardioverter-defibrillators reduces shocks in primary prevention patients: Results from the PREPARE (Primary Prevention Parameters Evaluation) study. J Am Coll Cardiol 2008; 52:541-550.



Your comments
Routine shock-energy testing in primary-prevention ICDs questioned—again
# 1 of 1
August 12, 2008 03:46 (EDT)
F. Roosevelt Gilliam
If only 97% right were good enough
What should be noted is the great success that all ICD units have at implant. The testing identifies few patients that will require more electrode surface area ( array, etc) or often placement of the lead in a better location. In addition to evaluating DFT the integrity of the system is verified. While new systems have impedence measurements, they may fail to note certain fractures until high voltage is passed across the system.
It is the rare patient whowill benefit and the numbers in these studies do not have the power to find such small numbers. this is very different from saying this is not important

You have to be logged in to add a comment to this article
Login
Username 
Password 
  Forgot your password?
 
Remember me on this computer
 
Join theheart.org community
Five reasons to become a member of the most trusted source of cardiology news:
1Be part of the conversation in our blogs and discussion forum
2Share your thoughts on our news or educational programs
3Receive exclusive newsletters related to your field of interest
4Access unique continuous medical education content
5See and read what leaders have to say about cardiology today
It is free and it only takes five minutes to join!
 
button
Latest 5 articles from Arrhythmia/EP
Previews
Featured CME
Inside: Arrhythmia/EP